Lap Laser Dorado-3 System

Moving Laser System for Patient Alignment

Used in Radiation Therapy

lap Laser Dorado-3 multi-laser patient positioning system, used in radiation therapy

The Dorado-3 is ideal for precise patient marking, accurate planning and exact positioning. Patient marking takes place during CT simulation (virtual simulation) and is required for reproducible treatment positioning on the LINAC. The Dorado-3 Laser System consists of one ceiling and two wall post rails with fixed and movable red laser modules to mark the patient in all three body planes.

The Dorado-3 includes

Options, sold separately

Notes:

Dorado-3 Mounting Options

wall-ceiling-call configuration for Dorado-1 laser system post-ceiling-post configuration for Dorado-1 laser system post-ceiling-wall configuration for Dorado-1 laser system wall-ceiling-post configuration for Dorado-1 laser system bridge configuration for Dorado-1 laser system

Dorado-3 Laser System Specifications

Dorado-3 Laser Specifications
System
Dimensions Wall
Post
Bridge
1553 × 143 × 109.5 mm
1707.5 × 225 × 164 mm
2594-5000 mm x 2300-2800 mm
Weight Wall
Post
Bridge
23 kg
24 kg
approx. 100 kg
International Protection Rating IP20
Operating temperature 15 - 30°C
Ambient conditions 35 - 80 % relative humidity, non-condensing
Travel range (ceiling) 700 mm
Travel speed (ceiling) up to 200 mm/s
Positioning accuracy ± 0.1 mm
Projection precision (ceiling) ± 0.5 mm up to 4 m distance
Laser
Laser color (typical wave length) red (638 nm)
Laser class 2
Focusable range 1 - 4 m
Line length at 3 m distance > 3 m
Line width up to 4 m distance < 1 mm
Maximum laser output power < 1 mW
Power Supply
External power supply 100 - 240 V AC, 50 - 60 Hz
Internal voltage 24 V D
International Certifications ISO-9001, ISO-13485